You are here:
Cardiac Science recalls the Powerheart AED model 9200RD und 9212RD
2005.02.01
Product group Medical electronics / Electromedical devices
Reference 02513/04
Cardiac Science has initiated a recall of the Powerheart AED model 9200RD und 9212RD that were shipped to customers in the late 2002 and early 2003. A marginal capacitor component may cause a longer than specified charge time and lower than specified energie output.
Cardiac Science recalls the Powerheart AED model 9200RD und 9212RD , Download_VeroeffentlichtAm_EN